Is there a tool for the diagnostic test of the probe?

icon icon
Yes, please open Apache app, click the drop-down list in the upper left corner and press “Help & About” for a diagnostic test.
icon icon

After powering on, the probe’s green light keeps blinking slowly. What does this mean?

icon icon
If multiple reboot attempts still result in the same issue, please contact your local distributor to use the Recovery Tool for restoration or send the probe back to the manufacturer for system recovery.
icon icon

What should I do if the probe shows an orange light and won’t turn on?

icon icon
Please charge the device continuously for 6-8 hours and then attempt to power it on again. If it still fails to start, please contact your local distributor.
icon icon

How long do I need to charge my probe for?

icon icon
C62 / L154: System and charging dock approximately 2.5 hours.
neo series: Approximately 80 mins (5V, 2A).
icon icon

Can the probe be used until the battery is completely drained? At what battery percentage will it automatically shut down?

icon icon
The probe will automatically shut down when the battery is at 5% to protect battery health. It is recommended to avoid prolonged low-battery usage to maintain battery performance.
icon icon

Why does my probe update fail?

icon icon
1. Ensure the connection mode in the settings is not switched to SSID mode. If it is, disable this mode.
2. The probe must have a battery level of more than 20% to proceed with updates.
icon icon

What are the available clinical applications?

icon icon

For clinical applications and suitable probe options, please refer to the Specialities page.

Learn more about how to set presets

icon icon

What are the frequency and depth of the probe?

icon icon

L154 Frequency: 4~15 MHz, Depth: Max.12cm / C62 Frequency:  2~6 MHz Depth: Max. 30cm.

See Product page for more information.

icon icon

How to load patient’s information?

icon icon

User can enter patient information manually or load it from the "Worklist" on the bottom of the patient information page.

Learn how to set patient's information

icon icon

How to change the TGC on Apache?

icon icon
In the exam page, press floating toolbar for adjustments.