What happens if the probe gets warm?

icon icon
It needs some time to cool down. When the temperature gets warm after using for a while, the system has a temperature control for safety purposes. Please let it cool down for 10~15 minutes by freezing or turn off the probe before start using again.
icon icon

What is the battery life of the transducers?

icon icon

C62 / L154 : 2~2.5 hours for continuous scan (full charged).
neo Series : >= 80 minutes for typical scan

See Product Page for more information.

icon icon

Is there a way to know when the probe in low on battery?

icon icon
Yes, the percentage of battery shows on the upper-right corner in the scan page.
icon icon

How to turn on the probe?

icon icon

Simply press down and hold the power button for 1 second and refresh user's tablet in the probe list for connections.

See LED indications for more information.

icon icon

Is there a tool for the diagnostic test of the probe?

icon icon
Yes, please open Apache app, click the drop-down list in the upper left corner and press “Help & About” for a diagnostic test.
icon icon

After powering on, the probe’s green light keeps blinking slowly. What does this mean?

icon icon
If multiple reboot attempts still result in the same issue, please contact your local distributor to use the Recovery Tool for restoration or send the probe back to the manufacturer for system recovery.
icon icon

What should I do if the probe shows an orange light and won’t turn on?

icon icon
Please charge the device continuously for 6-8 hours and then attempt to power it on again. If it still fails to start, please contact your local distributor.
icon icon

How long do I need to charge my probe for?

icon icon
C62 / L154: System and charging dock approximately 2.5 hours.
neo series: Approximately 80 mins (5V, 2A).
icon icon

Can the probe be used until the battery is completely drained? At what battery percentage will it automatically shut down?

icon icon
The probe will automatically shut down when the battery is at 5% to protect battery health. It is recommended to avoid prolonged low-battery usage to maintain battery performance.
icon icon

Why does my probe update fail?

icon icon
1. Ensure the connection mode in the settings is not switched to SSID mode. If it is, disable this mode.
2. The probe must have a battery level of more than 20% to proceed with updates.